A fuzzy system cardio pulmonary bypass rotary blood pump controller

作者:

Highlights:

摘要

Fuzzy logic regulation of flow was investigated in simulation of a model of the Cleveland Clinic Foundation rotodynamic Cardio Pulmonary Bypass blood pump comprised of impeller model #4079 and volute model #4080. A non-linear model of the pump was derived from pump maps obtained from measurements using a custom made pump dynamometer. Pump speed and generated delta pressure were used as inputs to a fuzzy engine in charge of manipulating the pump's speed; the fuzzy controller was capable of maintaining a setpoint of 6±0.5 l/min when presented with pressure disturbances over a range of ±50 mmHg from the baseline of 100 mmHg.

论文关键词:Rotary blood pump,Fuzzy control,Non-linear model,Flow estimation

论文评审过程:Available online 20 November 2003.

论文官网地址:https://doi.org/10.1016/j.eswa.2003.09.006